Nuclear magnetic resonance spectroscopy, Nuclear magnetic resonance , NMR, NMR spectroscopy, Magnetic resonance imaging, MRI
MRI-300 Magnetic Resonance Imaging
Introduction
MRI-300 Magnetic Resonance Imaging MRI is specially designed for living animal’s observation. The MRI-300 Magnetic Resonance Imaging provides high contrast images and accurate messages in vivo intuitively, equipped with permanent magnet. As a powerful and non-destructive Magnetic Resonance Imaging(MRI) instrument, the MRI-300 Magnetic Resonance Imaging is widely used in life sciences to study tissue structure in vivo and their contrast message dispense

Applications
  • Nuclear magnetic imaging related research
  • Magnetic resonance imaging experiments, 2D/3D imaging research
  • Imaging research with living sample (mice, plants and etc.)
Specifications
  • Magnet: permanent,1.45±0.05T:
  • H resonant frequency: 61.7±2MHz
  • Magnet Pole diameter: ϕ200mm
  • Sample field size: ϕ45mm×H80mm
  • Magnet field uniformity:< 8ppm
  • Magnet temperature control: dual system of nonlinear precise temperature control, temperature controllable from 25°C to 35°C ; temperature control accuracy ±0.005°C ; temperature display accuracy 0.001°C
  • Signal to noise ratio (SNR): 55dB, Degree of distortion: <1%
  • Graphics resolution: max. 512x512x128(3D)
  • Magnetic field stability: Larmor frequency variations 100Hz/Hr
  • Space resolution: 0.15mm, max. 0.08mm
  • Gradient field Intensity: 1.1 mT/m
  • Max gradient magnet field: 200mT/m.s
  • Graphics linearity: >98% in X, Y and Z direction
  • Absolute resolution: <0.08mm
  • Power supply: 110V/220V, 60/50Hz
  • Work temperature: 10~35°C
  • Environment humidity: 30~70%
  • Dimensions (L, W, H): Main unit 500mm×500mm×700mm.