ICP, ICP-OES, ICP-AES, AAS, GC-MS, GC, HPLC, Inductively coupled plasma mass spectrometry, Inductively coupled plasma atomic emission spectroscopy, Inductively coupled plasma optical emission spectrometry, Atomic Absorption Spectrometer, Mass Spectrometer, High-performance liquid chromatography ,Gas Chromatography
LC500 HPLC
Introduction
LC500 HPLC features fully automatic digital computer control. Its electronic circuit design, internal mechanical structure design, processing technology, functions of cinematography workstation and the technical criteria make it a leading products with excellent stability and reliability.

Features
  • Fully automated digital intelligent computer control
  • In-parallel connected pumps with reciprocating double plungers improves the stability, reliability and the service life.
  • Noval appearance, convenient and simple operation and ergonomic design
  • Precise positioning structure in the optical path system
  • Heat insulation installation technology guarantees higher precision, smaller excursion, and shorter period for stabilization
  • All of components are made by leading digital processing centers of the world
Specifications - Pumps
Delivery method double pistons, main & auxiliary
Pump head options 10ml SS/PEEK analytical
50ml SS/PEEK semi-preparative
Maximum working presure 10ml pump head: 42Mpa for SS / 30Mpa for PEEK
50ml pump head: 30Mpa for SS / 20Mpa for PEEK
Flow Range 10ml pump head: 0-10 ml/min
50ml pump head: 0-50 ml/min
Flow rate accuracy RSD ≤ 0.5% (measurement at 1ml/min)
Flow rate precision RSD ≤ 0.1% (measurement at 1ml/min)
Specifications - UV detector
Wavelength Range 190 ~ 700nm
Baseline Noise Dynamic flow rate 1ml/min methanol/water=80/20, Wavelength 254 nm, < 4×10-4 AU/h
Baseline Drift Dynamic flow rate 1ml/min methanol/water=80/20, Wavelength 254 nm
Wavelength Accuracy ±2 nm
Wavelength Reproduction 0.4nm
Detection Range 0-2AU, <1.5 x 10-5 AU/h